CLOSURE OF LARGE PATENT DUCTUS ARTERIOSUS BY POST INFARCT VSD CLOSURE DEVICE: A CASE REPORT

dc.contributor.authorHussain, Ijaz
dc.contributor.authorZeb, Shah
dc.contributor.authorKhattak, Kamran
dc.date.accessioned2026-09-05T09:31:38Z
dc.date.issued2016-09-10
dc.description.abstractPatent ductusarteriosus (PDA) constitutes 6 to 11% of all congenital heartdefects. While surgical ligation and video-assisted, thoracoscopic interruption ofPDA are still available, transcatheterocclusion of PDAs has recently assumed amajor role in closure of PDA. Most of the centreuse Amplatzer duct occluderdevice for PDA closure.Here we report a large PDA that was not amenable toclose by largest Amplatzer duct occluder, so we used successfully 24mmAmplatzerpost infarct VSD occluder. Patient discharged home with stablehemodynamics and without complication as assessed by echocardiography 24hours after procedure.
dc.identifier.citationPakistan Heart Journal; Vol. 48 No. 3 (2015)
dc.identifier.doi10.47144/phj.v48i3.964
dc.identifier.urihttps://pakheartjournal.com/index.php/pk/article/view/964
dc.identifier.urihttps://ds.pakheartjournal.com/handle/phj/1494
dc.language.isoen
dc.publisherPakistan Cardiac Society
dc.relation.ispartofseriesPakistan Heart Journal; Vol. 48 No. 3 (2015)
dc.titleCLOSURE OF LARGE PATENT DUCTUS ARTERIOSUS BY POST INFARCT VSD CLOSURE DEVICE: A CASE REPORT
dc.typeArticle

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
964-1-1981-1-10-20151231.pdf
Size:
73.58 KB
Format:
Adobe Portable Document Format

Collections